GC Inlet Pressure Shutdown

What is the number one cause of an inlet leak?  The septum no longer sealing the inlet.  This is the very first thing I would change if my inlet could no longer maintain pressure.  This holds true for 5890, 6890, and 7890 HP/Agilent GCs.  Even if you just changed it, I would recommend changing it again just to be sure.  Think about it, this septum is actually being pierced by the needle over and over again, its only a matter of time before it will start leaking.

5183_4757

Another easy fix would be to tighten the column nut in the inlet.  This is located inside the oven on the left side.  As the GC heats up and cools down, its very possible for the nut to become loose.  When you tighten this, you want it finger tight (make sure its not hot!) and then about a quarter turn or a little more with a wrench should be good.

medium

Lastly you want to replace the liner and the o-ring if you are still having pressure issues.  Sometimes the glass liner can crack, or the if the o-ring is old it won’t make a good seal anymore.  Again, make sure the inlet is cool before touching anything.

Liners

These should take care of any minor leaks you may have, i.e. can build some pressure just can’t reach setpoint.

Now what if you can’t build any pressure at all?  In this case the first thing I would do is check that the supply gas is making it to the GC.  To do this slowly unscrew the gas line connection to the back of the GC (obviously you want to make sure you select the carrier gas line) and as soon as you start to loosen it you should hear ‘hissing’.  If not, then your problem is not with the GC, but with the supply gas.  If the supply  gas is fine, and you have 0 pressure reading on the inlet, I would suspect a more serious problem.  This could be the EPC module that is bad, or the inlet weldment could have a crack, same with the top insert assembly.  Further diagnosis can be achieved with an electronic leak detector, or a soapy solution such as snoop.

Another way to narrow this down:  You can open the top insert as if you were removing the liner, and just cover the bottom of that cap (where the two stainless steel tubes go) with a thumb or septum.  Turn the pressure on for the inlet and see if its able to build pressure.  If yes, the leak is downstream from there, inlet weldment or column nut.  If not, then the leak is before that, the stainless steel tubing or the EPC module.

High Pressure in my HPLC!

Ok so no one wants to see the pressure on their HPLC higher than what it should be.  In my experience the number one cause of high pressure is a clog in the autosampler.  Followed by a clog in the purge valve frit, as the second highest cause of HPLC over pressure.  Luckily, there’s an easy way to diagnose whether or not there’s a clog in the autosampler.  This technique works on Agilent model HPLCs and Chemstation, not sure about other brands.

Turn the pump on, once you see the high pressure, right-click on the diagram of the autosampler.  Select the switch valve to bypass option (see below).  See if pressure drops immediately, if so change the needle and needle seat (or backflush).

bypass

Note: Older versions of chemstation will only have that option available in the diagnosis tab.  I may make a tutorial on the process in the future.

What if the pressure is still high?  Try removing the column from the flow path.  If that doesn’t help, try removing the autosampler from the flow path.  If that doesn’t work, try opening the purge valve.  If opening the purge valve does not decrease the pressure, try replacing the purge valve frit.

Pro Tip:  When installing a brand new column, make note of the pressure of the system.  This is what your pressure baseline is, and will be a good reference in the future when you are not sure if your pressure is where it should be.

Chemstation Calibration Units

Pro Tip: Chemstation does not look at the units when calculating amounts.

Example:

amt

As you can see the amounts is in ng/ul.  If I go to calibration–> calibration settings, I can choose any unit (or non unit for that matter) I like.

amt2

In this example I chose Aliens as my unit of choice.

amt3

As you can see, Chemstation willfully changes the units to that value.

Why is this important?  Many people think that Chemstation will convert units for you.  You can add a dilution or multiplication factor, but that will be used in the calculation of the amount regardless  of what units you choose.  Best practice would be to build your calibration table based on the units which you would like your results to be, so that by the time you print the report, you won’t need to convert any units.

Blog at WordPress.com.

Up ↑